/tr>
Modes |
Assist – Volume Control
Assist – Pressure Control
Synchronized intermittent mandatory Ventilation (SIMV) – Volume Control
Synchronized intermittent mandatory Ventilation (SIMV) – Pressure Control
CPAP / PSV
BiPAP
|
FiO2(Fraction of Inspired O2 concentration) |
( Fraction of Inspired O2 concentration)Provision to connect
O2 supply. FiO2 % to be controlled externally
|
Respiratory Rate: |
0-60 bpm |
PEEP |
0-30 cmH2O
(increment of 0.5cmH2O) |
I:E ratio |
1:4 to 4:1
(increment of 0.1) |
Exhaled volume measurement |
YES |
Power |
Single phase 220VAC/110VAC 50/60Hz
Battery : Provision to connect 24V DC External Battery.
Seamless changeover between power source |
Exhaled air virus filter included
(Electrostatic filter HME)
|
YES
|
Microprocessor based system
|
Onboard Display : IP66, 7 inch Multicolor backlit, Graphical Display
Communication : USB port, Ethernet (TCP/IP), RS232/485
Expandable Memory for Data logging : 4GB to 32GB |
Configurable Alarms with
Audio Visual Indication |
Auto/Manual Reset, Action on Alarm
Self
Power supply
Peak Inspirator Pressure
Low Inspiratory Pressure (Airway disconnection)
PEEP > Set PEEP for n Consecutive cycles
VTi < Set VT 0.75 for 6 consecutive cycles
Respiratory Rate
System Faults: Memory, Temperature
FiO2 < Set FiO2 for n consecutive cycl/es
SPO2 < Set SPO2 value (Normally 90%) (Optiona/l)
Vital components inactive |
Graphics mode for Display
and monitoring of Set and
Actual parameters |
Set parameters: Respiration Rate, Tidal Volume, PEEP, I:E,
Pressures, Flow, Apnea, FiO2 etc.
Actual parameters: Scalars (Volume/Pressure/Flow Vs time),
Loops (Flow/Pressure Vs Volume),